Ventllogic LS & Plus – Rental
100 % Mobility and Reliability in IV and NIV
Your requirements for reliability and mobility are our benchmark.
VENTI logic LS and VENTI logic plus are the forerunners in the new generation of venti-lators. They offer you a high degree of reliability and versatility every day at all times. Their practice-oriented monitoring and mobility concepts are supplemented by unique ventilation functions. VENTI logic LS and VENTI logic plus have leakage and single patient circuits. In addition, VENTI logic LS offers a double patient circuit system with patient valves and volume-controlled ventilation modes (VCV, aVCV).

Technical Data Ventllogic LS & Plus
Product class as per Directive 93/42/EEC: | llb |
Dimensions (W x H x D): | 240 x 153 x 340 mm |
Weight | |
without replaceable battery: | about 5.9 kg |
with replaceable battery: | about 6.5 kg |
Temperature range | |
Operation: | +5 °C to +35 °C |
Storage: | -40 °C to +70 °C |
Air pressure range: | 600 – 1100 hPa (below 700 hPa leakage is to be kept low because the device may not be able to compensate for high ventilation pressures) |
Electrical connections: | 115 – 230 V AC, 50 – 60 Hz Tolerance -20 %, +10 % |
Power consumption at | 230V 115V |
Operation: | 0,35 A 0,70 A |
Standby: | 0,05 A 0,10 A |
Maximum power consumption: | 120 W |
Switching capacity | |
Remote alarm connection: | 60 V DC/2 A; 42 V AC/2 A |
Battery capacity | |
internal rechargeable battery: | 4.5 hours |
replaceable/rechargeable battery: | 4.5 hours |
*) The capacity depends on the ventilation parameter settings and the battery‘s age and state of charge. | |
*)Classification as per EN 60601-1 | |
Protection from electric shock: | Protection class II |
Degree of protection from electric shock: | Type BF |
Time required to charge battery: | |
Charge via ventilator: | about 6 hours per battery |
Leakage modes in both devices: | CPAP, S, ST, T, TA |
and only VENTI logic LS: | SX, SXX |
Valve ventilation modes in both devices: | PSV, PCV, aPCV, SIMV |
and only VENTI logic LS: | VCV, aVCV |
and only VENTI logic LS: | VCV, aVCV |
Special therapeutic functions: | |
AirTrap Control | LIAM |
Trigger lockout | Volume compensation |
Expiratory pressure ramp | three ventilation programs |
Electromagnetic Compatibility | |
Radio interference suppression: | EN 55011 |
Radio interference resistance | EN 61000-3-2, EN 61000-3-3, EN 61000-4-2 to 6, EN 61000-4-8, EN 61000-4-11 |
Mean sound level / operation as per EN ISO 17510 with 1 m distance between device and patient position: | about 28 dB(A) at 10 hPa |
Sound level of alarm: | about 69 dB(A) as per EN 60601-1-8 |
IPAP pressure range: | 6 to 40 hPa (leakage circuit) 4 to 40 hPa (valve system) |
PEEP/EPAP | 4 to 20 hPa (leakage circuit) |
pressure range: | 0 to 20 hPa (valve system) |
CPAP pressure range: | 4 to 20 hPa (leakage circuit) |
Pressure accuracy: | to 35 hPa ± 0.8 hPa from 35 hPa ± 1.5 hPa |
Increment: | 0.2 hPa (1 hPa = 1 mbar ≈ 1 cm H 2 O) |
Tidal volume: | 50 – 3000 ml |
Minimum pressure limit stability (PLSmin) (min. pressure in case of device failure): | ≥ 0 hPa |
Minimum pressure limit stability (PLSmax) (max. pressure in case of device failure): | ≤ 60 hPa |
Respiratory rate: | 5 to 45 bpm |
Accuracy: | ± 0.2 bpm |
Increment: | 0.5 bpm |
I:E-ratio | |
Inspiration time: | 15 % to 67 % of breathing period |
Increment: | 1 % |
Accuracy: | ±1 % |
Trigger level: | adjustable in 8 stages for inspiration and 14 stages for exhalation (from 5 % to 95 % of maximum flow), can be switched off for exhalation in ST mode |
Pressure increase | |
speed: | Can be set in 6 levels |
Pressure decrease speed | |
Leakage system: | Can be set in 6 levels |
Valve system: | One permanently set level |
Accuracy | |
Volume measurement: | at 23 °C: ±20 %, at least 25 ml |
Max. allowable flow | |
with oxygen feed: | 15 l / at ≤ 1000 hPa |
Max. heating of respiratory air at 35°C ambient temperature: | 41°C |
Pressure constancy measured as per DIN EN ISO 17510 in CPAP mode: | < 10 hPa: ∆p ≤ 0.5 hPa |
DIN EN ISO 17510 in CPAP mode: | > 10 hPa: ∆p ≤ 1.0 hPa |
Fine filter separation level to 2 um: | >≤ 99.7 % |
Fine filter service life: | >1000 hours in normal ambient air |
Flow at max. speed at 0 hPa: | Leakage ventilation: 350 l/min Single patient circuit with patient valve: 345 l/min Double patient circuit with patient valve (only VENTI logic LS): 345 l/min ±15 l/min |
Tolerance: | ± 15 l / min |